Specially designed for sterilization by means of 100% Ethylene Oxide, working with sub-atmospheric technique or an anti-explosion combination of ethylene oxide with inactive gas.
Internal chamber, double walls and doors comply with the ASME code, and are completely made of type 316 L stainless steel. The inner surfaces in contact with gas are sanitary polished. The chamber is thermally insulated.
   
Depending on the model, sliding doors can be vertical or horizontal. They close airtight by means of a pressurized silicone seal inserted in the autoclave.
Sterilizers can be single or double door. Double door equipment is fitted with safety systems which prevent both doors from opening simultaneously.
The equipment is fitted with a microprocessed control with Touch Screen panel. It is user friendly making it possible to visualize and control all functions from a single screen. It can store 99 cycles, completely programmable by means of a password, ensuring absolute security in the operation. It has a parallel alarm program which verifies the process parameters.
   
It can store up to 99 cycles in its memory. There are seven programs with direct access for those cycles more frequently used. All cycles may be modified according to the requirements of the processes to perform.
Before starting a cycle, the operator may enter the data of the product to be sterilized so that they are printed in the cycle graphic record, enabling the correct traceability of the product.
Sterilizers are anti-explosion in case ethylene oxide is used. Equipment complies with safety norms to avoid an explosion.
An alphanumerical printer records all cycle stages and variables. All the information needed to ensure the correct fulfillment of the cycle parameters is recorded.
Loading system consisting of internal and external carts for loading and carrying the materials to be sterilized in the autoclave. The internal cart is made of type 316 stainless steel with lateral runners for trays carrying materials, adjustable in height, and lower wheels made of type 316 stainless steel. The external cart is made of type 304 stainless steel, with four tyred wheels, two fixed and two spinning. It is fitted with upper rails made of type 304 stainless steel with leveling and fixing screws.

It neutralizes ethylene oxide at the outlet of the equipment after degasification. It is installed at the effluent outlet and it operates by means of hydrolysis in an acidified water flow.
Weight system for ethylene oxide tubes which enables gas dosage for each cycle and control of gas excessive consumption or leak. The system records the initial and final weight of the tube after each cycle. It consists of two scales. When one tube runs out of ethylene oxide, it automatically switches to the other without stopping the cycle. The system prevents the beginning of the cycle if the amount of ethylene oxide is not enough for the sterilization process.
     
Special equipment with chamber capacity for more than 20.000 liters available upon request. We reserve the right to make alterations without notice

Documentation: Before delivery, our equipment is properly certified in our plant performing the corresponding Factory Acceptance Test (F.A.T). Once installed, the Site Acceptance Test (S.A.T) is performed.

Thermal Profiles: All instruments used are traceable to primary patterns to measure the temperature of a process, or to verify control or measuring instruments.

 